| Specifications |
models |
| |
391A |
390A |
389B |
388B |
| Volts |
True RMS reading |
|
| DC Ranges |
200 mV, 2V, 20V, 200V, 1000V |
400 mV, 4V,
40V, 400V, 1000V |
| AC Ranges |
200 mV, 2V, 20V, 200V, 750V |
400 mV, 4V,
40V, 400V, 750V |
| Resolution |
10uV, 100mV, 1mV 10 mV,
100 mV |
100uV, 1 mV,
10 mV, 100 mV, 1 V |
| DC Accuracy |
+-(0.05% + 3 dgt) |
+-(0.1% rdg + 2 dgt) |
+-(0.25% rdg + 2 dgt) |
+-(0.5% rdg + 1 dgt) |
| AC Accuracy |
+-(1%+10 dgt.) 50 Hz - 500
Hz |
400 mV: +- (1.2%
rdg + 5 dgt) |
+-(1.2%rdg + 4 dgt) |
| |
+-(2% + 10 dgt.) 500 Hz
- 2 kHz |
50 Hz - 100
Hz |
750V: +- (2% rdg + 4 dgt) |
| |
500 Hz on 750V range |
+-(1.0% rdg + 3 dgt) 50
Hz - 500 Hz |
>10% of range at 200mV range |
+-(1.5% rdg + 3 dgt) 500
Hz-1kHz |
| Overload Protection |
1200VDC or AC rms |
1100 VDC or
AC rms |
1200VDC or AC rms |
| |
500VDC/AC rms 15 sec |
|
|
500VDC/AC rms 15 sec |
| |
on 200mV range |
|
|
on 200mV range |
| Input Impedance |
10 MOHM |
400mV: > 100MOHM,
4V: 10 MOhm |
40V - 1000V: 9.1MOHM |
10MOHM |
| Current |
True RMS reading |
|
|
| Ranges |
200 uA, 2 mA, 20 mA, 200
mA, 20A* |
400 uA, 4 mA,
40 mA, 400 mA, 20A* |
400 mA, 4 mA, 40 mA, 400
mA, 2A, 20A* |
| Resolution |
10nA, 100 nA, 1 uA, 10 uA,
1 mA |
0.1 uA, 1 uA,
10 uA, 100 uA, 10 uA |
100 nA, 1 uA, 10 uA, 100
uA, 1 mA, 10 mA |
| DC Accuracy |
200 uA to 200 mA: +-(0.5%
rdg + 5 dgt.) 20 A: +-(2% rdg + 10 dgt.) |
400 uA - 400 mA: +-(1% rdg
+ 5 dgt) 20 A: +- (2% rdg + 3 dgt) |
+-(1% rdg + 1 dgt)
+-(2% rdg + 3 dgt) |
400 uA - 400 mA: +-(1% rdg
+ 1 dgt)
2A: +-(1.5% rdg + 1 dgt)
20A: +-(3% rdg + 3 dgt) |
| AC Accuracy |
200 uA to 200 mA: +-(1.2%
rdg + 10 dgt.)
20A: +-(2.5% rdg + 20 dgt.) |
400 uA to 400
mA: +-(1.5% rdg + 4 dgt)
20A: +-(2.5% rdg + 4 dgt.) |
400 uA - 400 mA: +-(1.5%
rdg + 1 dgt)
2A: +-(2% rdg + 4 dgt)
20A: +-(3.5% rdg + 4 dgt) |
| Input Protection |
uA/mA Input:
0.5A/500V fast blow ceramic
fuse
20A input:
20A/600A fast blow ceramic
fuse |
uA/mA Input:
0.5A/500V fast blow ceramic
fuse
20A input:
20A/600A fast blow ceramic
fuse |
uA/mA input:
2A/600V fastblow ceramic
fuse
20A input:
20A/600V fastblow ceramic
fuse |
| Max. Burden Voltage |
600mV (900 mV on 20A range) |
500mV (2V on
4mA, 400mA ranges) |
600mV (900 mV on 2A, 20A
ranges) |
| Resistance |
|
|
| Ranges |
200OHM,
2kOhm, 20 kOhm, 200 kOhm, 2 MOhm, 20 MOhm |
400 OHM,
4 kOhm, 40 kOhm, 400 kOhm, 4 MOhm, 40 MOhm |
| Resolution |
10 mOHM,
100 mOhm, 1 Ohm, 10 Ohm, 100 Ohm, 1 kOhm |
100 mOHM,
1 Ohm, 10 Ohm, 100 Ohm, 1 kOhm, 10 kOhm |
| Accuracy |
200 OHM,
2 MOhm: +-(0.25% rdg + 10 dgt.)
2 kOHM
to 200 kOhm: +-(0.15% rdg + 3 dgt.)
20 MOHM:
+-(1.0% + 10 dgt.) |
400 OHM
to 400 kOhm: +-(0.5% rdg + 4 dgt.)
4 MOHM:
+-(1% rdg + 5 dgt.)
40 MOHM:
+-(2% rdg + 5 dgt.) |
400 OHM:
+-(1% rdg + 4 dgt.)
4 kOHM
to 4 MOhm: +-(0.8% rdg + 4 dgt.)
40 MOHM:
+-(2% rdg + 5 dgt.) |
| Open Circuit Voltage |
3.2 VDC typ. |
-0.45 VDC typ.
(-1.2 VDC on 400 OHM range) |
0.6 VDC typ. (3.2 VDC on
400 OHM range) |
| Overload Protection |
500 VDC or AC rms |
500 VDC or AC
rms |
|
| Diode Test |
Tested at 1 mA, 3.2 VDC
max. typ.
+-(1% rdg + 10 dgt.) accuracy |
Tested at 1.2
mA, 3.0 VDC max. typ.
+-(1.5% rdg + 3 dgt.) accuracy |
Tested at 1 mA, 3.2 VDC
max. typ.
+-(1.5% rdg + 3 dgt.) accuracy |
| Transistor Test (hFE) |
Does not apply |
Does not apply |
hFE range: 0 - 1000
hFE base current: 10 uA
DC |
| Capacitance |
|
|
|
| Ranges |
Does not apply |
4 nF, 40 nF,
400 nF, 4 uF, 40 uF, 400 uF, 4 mF, 40 mF |
4 nF, 40 nF, 400 nF, 4 uF,
40 uF |
| Resolution |
|
1 pF, 10 pF, 100 pF, 1 nF,
10 nF, 100 nF, 1 uF, 10 uF |
0.1 pF, 1 pF,
10 pF, 100 pF, 1 nF |
| Accuracy |
|
4 nF: +-(3%
rdg + 20 dgt.) 4 nF to 40 uF: +-(3% rdg + 5 dgt.)
400uF to 40 mF: +-(5% rdg
+ 10 dgt.) |
+-(3% rdg + 4 dgt.) |
| Test Voltage |
|
< 1V |
< 3.5V |
| Overload Protection |
|
500 VDC or AC
rms |
| Frequency |
|
|
|
| Ranges |
2 kHz, 20 kHz, 200 kHz |
4 kHz, 40 kHz,
400 kHz, 4 MHz, 40 MHz |
4 kHz, 40 kHz, 400 kHz,
4 MHz |
| Resolution |
0.1 Hz, 1 Hz, 10 Hz |
1 Hz, 10 Hz,
100 Hz, 10 kHz, 100 kHz |
1 Hz, 10 Hz, 100 Hz, 10
kHz |
| Accuracy |
+-(0.1% rdg + 3 dgt.) |
+-(0.1% rdg + 3 dgt.) |
+-(0.25% rdg + 4 dgt.) |
+-(0.1% rdg + 2 dgt.) |
| Sensitivity |
50 mV rms min. (@>30 &
<70% duty cycle: 400 mV rms min.) |
1 Hz - 4 MHz:
1V rms
4 MHz-40MHz: > 2V rms, <5V
rms |
250 V rms min. on 10 Hz
to 1 MHz
500 mV rms min. on 1 MHz
to 4 MHz |
| Minimum Pulse Width |
> 25 ns |
> 25 ns |
> 2 us |
| Duty Cycle Limits |
> 30% & <70% |
> 30% &
<70% |
| Minimum Input Range |
2 kHz: 10Hz: 20 kHz: > 60
dgt; 200 kHz > 60 dgt |
|
| Overload Protection |
500 VDC or AC rms |
500 VDC or AC
rms |
| Logic Test |
|
|
| Logic Threshold |
Hi=2.8+-0.8V, 0=0.8 +-0.5V |
Does not apply |
Does not apply |
Hi=2.8+-0.8V, 0=0.8 +-0.5V |
| Frequency Response |
20 MHz |
|
|
20 MHz |
| Pulse Width |
25 ns |
|
|
25 ns |
| Pulse Limits |
> 20% and < 80% |
|
|
> 20% and < 80% |
| Indication |
40m sec beep at logic 1
(Hi) |
|
|
40m sec beep at logic 1
(Hi) |
| Overload Protection |
500 V DC or AC rms |
|
|
500 V DC or AC rms |
| Temperature |
|
|
| Range, Resolution |
Does not apply |
-58 0 to +2372
0
F, 1 0 F (-50
0 to +1300 0 C) |
Does not apply |
Does not apply |
| Accuracy |
|
+-(0.8% rdg + 2 0
C) -50 0 - 400 0 C
+-(1% rdg + 2 0
C) 400 0 - 1300 0 C) |
|
|
| Duty Cycle |
|
|
| Range, Resolution |
0 to 90%, 0/1% |
Does not apply |
Does not apply |
Does not apply |
| Accuracy (5V logic) |
+-(2.0% + 10 dgt.) |
|
|
|
| Minimum Pulse Width |
10 us |
|
|
|
| Frequency Range |
40 Hz to 20 kHz |
|
|
|
| Overload Protection |
500 VDC or AC rms |
|
|
|
| General |
|
|
|
| Display |
20000 count, 4 1/2 digit
LCD |
4000 count,
3 3/4 digit LCD with 41 segment analog bar graph |
4000 count, 3 3/4 digit
LCD |
| Polarity |
Automatic, positive
implied, negative polarity indication |
| Operating Temperature |
32 0
to 122 0 (0 0 to 50 0 C), 0 to 70% R.H. |
| Dimensions (HxWxD) |
7.8x3.5x1.57''
(198x90x40 mm) |
| Weight |
11.3 oz. (320g) |
| Three Year
Warranty |
| Accessories |
SUPPLIED: 9V
Battery, Test Leads, Instruction Manual, Thermocouple Probe (Model 390)
OPTIONAL: TL-2A Deluxe Test
Leads, TL-3 Accessory Tip Kit (for TL-2A), PR-28A High Voltage Probe (40kVDC),
TL-130A General Purpose DMM Kit
Carrying Case (not included):
LC-29B |
| *10 A continuous,
20A for 30 seconds max. |